How to Calibrate a Digital Thermometer with a Dry-Block Calibrator and with Freezing and Boiling Water

This video demonstrates two methods for calibrating a digital thermometer: first, with ice and boiling water, and second, with our Fluke dry-block calibrators. Both methods are easy and fast. However, the water method has inaccuracies that are unknown caused by variations in altitude, atmospheric pressure, and water purity. The dry-block calibration method provides a better measurement because we’re working with instruments that have a known accuracy.

 

Video transcript:

This video demonstrates two methods for calibrating a digital thermometer: first, with ice and boiling water, and second, with our Fluke dry-block calibrators. Both methods are easy and fast. However, the water method has inaccuracies that are unknown caused by variations in altitude, atmospheric pressure, and water purity. The dry-block calibration method provides a better measurement because we’re working with instruments that have a known accuracy.

Let’s experiment with a couple different methods of calibration. We’re going to use the common methods here of ice water and boiling water. I’ve got everything set up here, and we’re going to do a quick check with these, compare against our dry wells, and see how we measure up.

I’ve got the unit under test sitting in an ice bath and right now you can see we’re measuring approximately 0.6 degrees, a little bit off from where we want to be. We’re going to make a quick adjustment to get this dialed in to our ice bath. After a quick adjustment, we’re right about zero degrees.

Now I’m going to move over and try this at the boiling point and see how we’re doing on the linear side going up. It looks like we’re bobbling around about 95 degrees and some change, so let’s make an adjustment on the top end.

You can see I’ve made the adjustment on the top end, getting it close to 100 degrees Celsius.

So now what I’d like to do is repeat this, but this time we’re going to use our 9100S and 9102S and see how we measure up…

Just a few seconds ago you saw me make the adjustments using the water method. Now you can see we’re a little bit out of spec again. Again, those common methods can net you some errors in the neighborhood of several degrees. The benefit of moving to a dry well is that these are calibrated, certified, and you know the uncertainty you are working with. Boiling water depends not only on elevation, but also barometric pressure. Ice water and building an ice bath, there is some technique to it, and the purity of the water will also make a difference.

So let’s make another adjustment to the thermometer, and see what our results are.

I’ve moved right over to the boiling point because that’s where we left off, and you can see right here from the boiling water, to the dry well, we’re almost four degrees and some change high. So you can see now I’ve made the high side adjustment, and again, we’ve pulled it back from the boiling water, and now I’ve got it down to 100 degrees Celsius. Next, we’re going to move to our zero point, and see how we measure up.

Alright, we’ve allowed a few minutes for the device under test to come down to zero and stabilize. We’re actually right on with what the dry well is reading. So again, we’ve had to make a slight adjustment of about four degrees at our top end versus the boiling water. At our low end, we’re pretty linear.

So we know at this point we’re within 0.25 degrees Celsius of our reading, which is well within a Type K uncertainty, and again, we’re confident even at the top end we’ve got an uncertainty of 0.25 degrees as well.

You can see the difference between using water versus a dry well. The big benefit is you’re working with a known uncertainty with dry wells, versus an unknown with water.

Keep learning

Do you use ice and boiling water to calibrate temperature? There's a better way. Video.
How to Calibrate an RTD or Platinum Resistance Thermometer (PRT) application note
How to Calibrate an RTD on-demand webinar
How to do a temperature sensor comparison calibration blog post
Thermocouple App Notes Series, 4 of 4 - How to Calibrate a Thermocouple application note

See related products

Platinum resistance thermometers (PRTs)
Dry well temperature calibrators and micro-baths
Dry block and other industrial calibrators
Calibration baths

Get help

Customer & technical support
Speak with a calibration product expert about your calibrator or thermometer needs

  • Home
  • Products
    • New Products
    • Electrical Calibration
      • Electrical Standards
      • Electrical Calibrators
      • Bench Multimeters
      • Electrical Calibration Software
    • RF Calibration
      • RF Reference Sources
      • RF Calibration Accessories
      • RF Calibration Software
    • Data Acquisition and Test Equipment
      • Bench Multimeters
      • Data Acquisition
      • Data Acquisition Software
    • Temperature Calibration
      • ITS-90 Temperature Standards
        • ITS-90 Fixed-point cells
        • Standard Platinum Resistance Thermometers
        • Maintenance Apparatus
        • Liquid Nitrogen Comparison Calibrator
        • Resistance Bridges
        • Standard Resistors
      • Calibration Baths
        • Compact Calibration Baths
        • Standard Calibration Baths
        • Special Application Baths
        • Bath Accessories
        • Bath Controllers
        • Bath Fluids
      • Industrial Calibrators
        • Field Metrology Wells
        • Metrology Wells
        • Handheld Calibrators
        • Field Dry-Block Calibrators
        • Micro Baths
        • Infrared Calibrators
        • Thermocouple Furnaces
        • Dual Block Dry-Well
        • Zero-point Dry-Well
      • Probes / Sensors
        • Platinum Resistance Thermometers (PRTs)
        • Thermistors
        • Thermocouples
      • Digital Thermometer Readouts
      • Multifunction Calibrators
      • Temperature Calibration Software
    • Humidity Calibration
      • Humidity Generators
      • Humidity Data Loggers and Monitors
    • Pressure Calibration
      • Piston Gauges
        • NMI Piston Gauges
        • Absolute Piston Gauges
        • High Pressure Pneumatic Piston Gauges
        • Hydraulic Piston Gauges
        • Piston Gauge Accessories
      • Automated Pressure Controller / Calibrators
        • Low Pressure Controllers / Calibrators
        • Pneumatic Pressure Controllers / Calibrators
        • High Pressure Pneumatic Controller / Calibrators
        • Hydraulic Controller / Calibrators
      • Pressure Monitors
        • Low Pressure Monitors
        • Digital Pressure Gauges
        • Reference Pressure Monitors
      • Deadweight Testers
        • Pneumatic Deadweight Testers
        • Oil Deadweight Testers
        • Water Deadweight Testers
        • High-Pressure Hydraulic Deadweight Testers
        • Deadweight Tester Accessories
      • Manual Pressure Calibration
        • Pressure Calibrators
        • Pneumatic Pressure Control
        • Hydraulic Pressure Comparators / Pumps
      • Handheld Pressure Calibrators
      • Air Data Calibration
      • Environmental Monitors
      • Pressure Calibration Accessories
      • Pressure Calibration Custom Systems
      • Pressure Calibration Software
    • Flow Calibration
      • Gas Flow Standards
      • Gas Flow Accessories
      • GFS Primary Gravimetric Flow Standard
      • Flow Calibration Software
    • Process Calibration Tools
      • Temperature Calibrators
        • Handheld Temperature Calibrators
        • Dry-Block Calibrators and Micro-Baths
        • Precision Digital Thermometers
        • Temperature Probes
        • Infrared Calibrators
        • Hygro Thermometer with Data Logging
      • Pressure Calibrators
        • Digital Pressure Calibrators
        • Handheld Pressure Calibrators
        • Deadweight Testers
        • Precision Digital Pressure Gauges
        • Calibration Hand Pumps
      • Multifunction Calibrators
      • mA Loop Calibrators
      • Process Calibration Software
    • Calibration Software
      • MET/CAL® Software
      • MET/CAL® Support
      • Asset Management Software
      • Temperature Calibration Software
      • Pressure Calibration Software
      • Flow Calibration Software
      • Mechanical / Dimensional Calibration Software
    • Service and Support
    • All Calibration Instruments
    • Handheld Test Tools
  • Purchase Info
    • Where to Buy
    • Request a Quote
    • Request a Demo
    • Request a Sales Consultation
    • Certified Pre-Owned Equipment
    • General Services Administration (GSA)
    • Financing Program
    • National Stock Numbers (NSNs)
    • Payment Options and Tax Info
  • News
    • Press Releases
    • Promotions
    • Industry Links
    • Metrology Salary Survey
  • Training and Events
    • Conferences and Exhibitions
    • Training Courses
    • User Group Meetings
    • "How To" Seminars
    • Web Seminars
      • Live Seminars
      • On-Demand Seminars
  • Literature and Education
    • Education Hub
    • About Calibration
    • Articles and Education
    • Product Literature
    • Product Manuals (User Guides)
    • Videos and Virtual Demos
    • Resource Centers
    • Blog
  • Service and Support
    • Service Request (RMA)
    • Service Plans
    • Technical Support
    • Knowledge Base
    • Accreditations
    • Authorized Service Centers
    • Calibration Certificates
    • Community Forum
    • My MET/SUPPORT
      • Activate
      • Procedures
      • Software
      • Technical Bulletins
      • Priority Support
      • Manuals
    • Product Manuals (User Guides)
    • Safety Data Sheets (SDS)
    • Safety, Service, and Product Notices
    • Software Downloads
    • Tools
      • PT100 Calculator
      • PT100 Table Generator
      • ITS-90 Reference Function Calculator
      • Thermocouple Table Voltage Calculator
      • Thermocouple Voltage to Temperature Calculator
  • About Us
    • Contact Us
    • Visit Us
    • Employment
    • Fluke Calibration
      • Hart Scientific Temperature Calibration
      • DH Instruments Pressure and Flow Calibration
      • Pressurements Pressure Calibration
      • Ruska Pressure Calibration
    • Fluke Companies
    • Integrity and Compliance
    • Frequently Asked Questions
    • Why buy Fluke Calibration ?
Home|Literature|Articles and Education|Temperature Calibration|Video|How to Calibrate a Digital Thermometer with a Dry-Block Calibrator and with Freezing and Boiling Water

Products

  • Product index
  • New products
  • Where to buy
  • Request a quote

Services

  • Calibration and repair
  • About Fluke Calibration
  • Service centers
  • Software downloads

Resources

  • Catalogs
  • Manuals
  • Web seminars
  • Subscribe to e-news
  • Supplier Handbook

Company

  • Contact us
  • About Fluke Calibration
  • Why choose us
  • Other Fluke companies
©1995-2023 Fluke Corporation
             

Secondary menu

  • Disclaimer
  • Privacy Statement
  • Terms of Use
  • Terms of Sale