A typical 3-point calibration of 4 tri-clamp sanitary sensors takes over 4 hours (259.45 minutes) using a 7103 Micro-Bath. The same calibration using a 7109A Portable Calibration Bath takes about 2 hours (130.0 minutes). That’s a time savings of more than 2 hours per 4-sensor batch!
This table presents the details.
7109A Calibration Time Savings Estimate | ||||||||
Calibrate four tri-clamp sensors at three temperature points (0, 70, and 125°C) | ||||||||
7103 Micro-Bath | ||||||||
Temperature | ||||||||
Setpoint | Ramp range | Speed to temperature | Ramp Speed (minutes/°C) |
Start | End | Span | Stabilization Time (minutes) |
Calibration Time (minutes) |
0°C | 23 to 0°C | 25°C to -25°C: 0.9 min/°C | 0.9 | 23°C | 0°C | 23°C | 60 * | 80.7 |
70°C | 0 to 70°C | 25°C to 100°C: 0.47 min/°C | 0.47 | 0°C | 70°C | 70°C | 60 * | 92.9 |
125°C | 70 to 125°C | 25°C to 100°C: 0.47 min/°C | 0.47 | 70°C | 125°C | 55°C | 60* | 85.85 |
Total batch calibration time | 259.45 |
7109A Portable Calibration Bath | ||||||||
Temperature | ||||||||
Setpoint | Ramp range | Speed to temperature | Ramp Speed (minutes/°C) |
Start | End | Span | Stabilization Time (minutes) |
Calibration Time (minutes) |
0°C | 23 to 0°C | 25°C to -25°C: 1.5 min/°C | 1.5 | 23°C | 0°C | 23°C | 10 | 44.5 |
70°C | 0 to 25°C | -25°C to 25°C: 0.7 min/°C | 0.7 | 0°C | 25°C | 25°C | n/a | 17.5 |
25 to 70°C | 25°C to 140°C: 0.48 min/°C | 0.48 | 25°C | 70°C | 45°C | 10 | 31.6 | |
125°C | 70 to 125°C | 25°C to 140°C: 0.48 min/°C | 0.48 | 70°C | 125°C | 55°C | 10 | 36.4 |
Total batch calibration time | 130.0 | |||||||
The 7109A saves over 2 hours per four sensor batch | Time savings | 129.45 | ||||||
*The 7103 Micro-Bath calibrates one tri-clamp sensor at a time. Each sensor takes 15 minutes to stabilize for a total of 60 minutes per batch of four sensors. |
Learn more about the 6109A and 7109A Portable Calibration Baths.