- Other Fluke companies:
- Fluke
- Fluke Biomedical
- Fluke Networks
- Fluke Process Instruments
A Comparison between a Resistance Bridge and an Integrated-Circuit Resistance Thermometer Readout Used for SPRT Calibration
Specially designed bridges are used for standard platinum resistance thermometer (SPRT) calibration to achieve lowest uncertainty. Such bridges are expensive and their measuring speeds are slow. A new type of instrument reported a few years ago was compared against the bridge to see if it is possible to use the new instrument instead of the bridge for SPRT calibration in some cases. Four SPRTs were calibrated at the triple point of water and the freezing points of tin and zinc using a Model 6010T Bridge and the new instrument (Model 1590) simultaneously. At these calibration points the maximum differences between the two instruments were within 0.4 mK at the tin point, and within 0.7 mK at the zinc point. The maximum difference in resistance ratio W(t) at these points was within 0.9 ppm of the readings. The differences over the entire range from 0°C to 419.527°C were calculated for the four SPRTs. The maximum differences were within 0.1 mK close to 0°C, within 0.5 mK at 300°C and within 0.7 mK at 420°C. The comparison results show the new instrument can be used for SPRT calibration to achieve an expanded uncertainty (k=2) as low as 1.5 mK.
- Home
- Products
- New Products
- Electrical Calibration
- RF Calibration
- Data Acquisition and Test Equipment
- Temperature Calibration
- Humidity Calibration
- Pressure Calibration
- Flow Calibration
- Process Calibration Tools
- Calibration Software
- Service and Support
- All Calibration Instruments
- Handheld Test Tools
- Purchase Info
- News
- Training and Events
- Literature and Education
- Service and Support
- About Us