Precision, low inductance shunts for dc and ac current metrology
The A40B dc and ac precision current shunt set consists of 14 low-inductance coaxial current shunts, adapters and connectors, and a rugged transit/storage case. Designed for currents from 1 mA to 100A, the current shunts are for laboratory use in making directly measured absolute ac or dc current measurements as well as ac/dc current transfer measurements. They exhibit outstanding resistance value stability, with an excellent self-heating power coefficient, and a low temperature coefficient. With all of these characteristics combined, the shunts offer direct measurement of current from dc to 100 kHz with higher accuracy than previous commercially available shunt technology. Now precision measurements can be made in a single step process rather than by more complex traditional ac/dc transfer methods.
- Simplifies calibration/verification of precision calibrators and current sources
- Shunts sized for currents from 1mA to 100A
- Usable from dc to 100 kHz
- 14 individual dc and ac current shunts with a 1, 2, 5 sequence over 6 decades of current
- Simple direct measurements, making ac/dc transfers unnecessary
- Stability typically better then ± 5.0 µΩ/Ω for one year
- Typical angular accuracy of better than ± 0.003º at 1 kHz
Application
The wide range of shunt values permit measuring from 0.1 mA to 100 Amps. The physical construction and the components used in the current shunts ensure that the frequency response is very flat (amplitude displacement error relative to dc resistance). Also the phase displacement at 100 kHz is small enough to be neglected in all but the highest accuracy measurements. Combined with excellent dc resistance stability, the shunts can be used to directly measure current through their full bandwidth. This simplifies precision ac current measurements as the complex ac/dc transfer measurement process will no longer be necessary for many precision current measurement applications. These characteristics make the A40B shunts ideal for traditional precision current applications such as calibrator verification. Also, the wide current range allows verification of high current transconductance amplifiers. The low phase shift error is critical for measurement of non-sinusoidal wave shapes, as found in power quality or sampling digital wattmeter measurement applications.
Technical overview
The shunts come in four different sizes. The full current shunt set includes:
- four fully enclosed dc and ac current shunts: 1 mA, 10 mA, 20 mA, 50 mA
- five small size radial dc and ac current shunts: 100 mA, 200 mA, 500 mA, 1 A, 2 A
- three medium size radial dc and ac current shunts: 5A, 10 A, 20 A
- two large size radial dc and ac current shunts: 50 A, 100 A
Voltage output is nominally 0.8 volts for the nominal rated current input. The output is measured by a voltmeter or detector, such as precision voltmeters, ac measurement standards, ac/dc transfer standards, or thermal voltage converters, fitting them to a wide variety of metrology applications.
The radial style dc and ac current shunt design gives high performance with minimum external magnetic fields. Plus, the open nature of the dc and ac current shunt physical design maximizes air flow so the shunts have a minimal power coefficient effects. This permits each shunt to be used over a wide range of currents with stable resistance characteristics.
The dc and ac current shunt resistance is designed to minimize interaction with the detector instrumentation. The 100 Amp shunt resistance is 8 milliohms, with shunt resistance increasing to 80 ohms for the 10 mA shunt. Additionally, the 1mA shunt includes an internal battery operated buffer amplifier. It drives the output voltage so the 800 ohm shunt resistance has minimal interaction with the measurement device.
Reports of calibration
The A40B series of precision DC and AC Current Shunts come with an ISO 17025 Accredited Certificate of Calibration.
Basic Electrical Specification Summary
Shunt Nominal Current |
Nominal Resistance (Ohms) |
Basic Specification ±μA/A, TCal ±1 °C, ≤50 % RH | ||||
DC | 1 kHz | 10 kHz | 30 kHz | 100 kHz | ||
1 mA | 800 | 20 | 55 | 75 | 75 | 150 |
10 mA | 80 | 20 | 26 | 26 | 26 | 26 |
20 mA | 40 | 20 | 26 | 26 | 26 | 26 |
50 mA | 16 | 20 | 23 | 23 | 23 | 23 |
100 mA | 8 | 20 | 24 | 24 | 24 | 24 |
200 mA | 4 | 20 | 26 | 26 | 26 | 26 |
500 mA | 1.6 | 21 | 27 | 27 | 27 | 28 |
1 A | 0.8 | 21 | 27 | 28 | 28 | 31 |
2 A |
0.4 | 21 | 27 | 30 | 30 | 48 |
5 A | 0.16 | 21 | 31 | 32 | 40 | 71 |
10 A | 0.08 | 26 | 37 | 60 | 61 | 92 |
20 A | 0.04 | 26 | 43 | 52 | 70 | 113 |
50 A | 0.016 | 32 | 55 | 80 | 81 | 144 |
100 A | 0.008 | 35 | 65 | 90 | 98 | 174 |
Maximum AC-DC Difference
Shunt Nominal Current: | Maximum AC-DC Difference (± ppm)1,2 | |
1 kHz | 100 kHz | |
1 mA3 | 53 | 150 |
10 mA | 20 | 40 |
20 mA | 18 | 30 |
50 mA | 13 | 16 |
100 mA | 14 | 27 |
200 mA | 17 | 28 |
500 mA | 17 | 21 |
1 A | 17 | 23 |
2 A | 17 | 44 |
5 A | 23 | 69 |
10 A | 28 | 98 |
20 A | 37 | 150 |
50 A | 47 | 180 |
100 A | 60 | 300 |
1 Specifications indicate the maximum flatness deviation from dc, and include both measured AC-DC difference and the uncertainty of measurement. They are stated at k=2, approximately 95 % confidence. 2 Includes 1-year stability of the AC-DC difference. 3 Specifications for the 1 mA current shunt are for TCal ± 1 °C. |
General Specifications | |||||||||||||||||||||||||||
Calibration Documentation | 17025 accredited report of calibration included | ||||||||||||||||||||||||||
Dimensions (maximum) |
|
||||||||||||||||||||||||||
Physical/mechanical parameters |
|
||||||||||||||||||||||||||
Operating environment |
|
||||||||||||||||||||||||||
Storage and transit environments (for models other than the 1 mA current shunt) |
|
Additional 1 mA current shunt specifications | |
Output resistance | 8 mΩ |
Maximum safe output current | 11 mA (e.g., 1 V output into 90 Ω) |
Maximum capacitive load | 800 pF |
Output voltage regulation | 15 ppm/100 pF |
Maximum output dc v offset | ± 100 μV (typical ± 25 μV) |
Typical error @ 1 MHz | < 2 % |
Battery specifications | |||||||||
Battery size | AAA (44.5 mm x 10.5 mm) | ||||||||
Battery technology | Nickel-Metal Hydride (NiMH) | ||||||||
Number of batteries required | 8 (in 2 groups of 4) | ||||||||
Nominal battery voltage | 1.2 V (4.8 V per group of 4) | ||||||||
Typical battery capacity | 800 mAh | ||||||||
Storage and transit environment to preserve the batteries |
|
||||||||
Maximum operating time between charges |
|
Model Name | Description | |
---|---|---|
A40B/SET |
Complete set of shunts including: |
|
A40B-001MA |
1 milliamp Current Shunt |
|
A40B-010MA |
10 milliamp Current Shunt |
|
A40B-020MA |
20 milliamp Current Shunt |
|
A40B-050MA |
50 milliamp Current Shunt |
|
A40B-100mA |
100 milliamp Current Shunt |
|
A40B-200mA |
200 milliamp Current Shunt |
|
A40B-500mA |
500 milliamp Current Shunt |
|
A40B-1A |
1 amp Current Shunt |
|
A40B-2A |
2 amp Current Shunt |
|
A40B-5A |
5 amp Current Shunt |
|
A40B-10A |
10 amp Current Shunt |
|
A40B-20A |
20 amp Current Shunt |
|
A40B-50A |
50 amp Current Shunt |
|
A40B-100A |
100 amp Current Shunt |
Accessories common to all models:
Accessory | Description |
---|---|
A40B-ADAPT/LC |
LC Male to LC Male adapter |
A40B-ADAPT/LCN |
Female to N Male inter-series adapter |
A40B-CAL/LC |
High current adapter to connect two shunts in series for measurement (LC to LC) |
A40B-CAL/N |
Low current adapter to connect two shunts in series for measurement (N to N) |
A40B-CASE |
Fiberglass Transit Case |
A40B-LEAD/4MM |
N to 4 mm double banana connector |
A40B-LEAD/N |
N Male to N Male lead |
A40B-ADAPT/SPADE |
LC male to spade lug terminals |
Product Manuals |
---|
Application Notes |
---|
Data Sheets |
---|
FAQs |
---|
Papers / Articles |
---|
Video |
---|
Fluke 5790B Measurement Standard Demo Video |
White Paper |
---|
- Home
- Products
- New Products
- Electrical Calibration
- RF Calibration
- Data Acquisition and Test Equipment
- Temperature Calibration
- Humidity Calibration
- Pressure Calibration
- Flow Calibration
- Process Calibration Tools
- Calibration Software
- Service and Support
- All Calibration Instruments
- Handheld Test Tools
- Purchase Info
- News
- Training and Events
- Literature and Education
- Service and Support
- About Us
Sidebar Request a Quote
Request a quote