Extra-deep wells for thermometry work requiring extra tank depth and ultimate stability
- Constant liquid levels through concentric-tube design
- Special design for sighting LIG thermometers
- Depth up to 60 cm (24 in)
- Optional interface packages control all settings
The models 6054, 6055 and 7007 have extra-deep wells for use with liquid-in-glass thermometers, SPRT calibrations, or other thermometry work requiring extra tank depth. They were originally designed for NIST.
Well depths vary from 17 to 24 inches to eliminate stem conduction effects in probes that require more than 12 inches of immersion. Originally developed for a national standards lab, these baths are optimized for the visual calibration of liquid-in-glass thermometers.
The 6054 covers the temperature range of 50 °C to 300 °C, is also 24 inches deep, and comes with a removable stainless steel cover.
The 6055 is engineered for the temperature range of 200 °C to 550 °C with salt and is 17 inches deep. Specific size differences and various specifications are shown in the comparison table.
The Model 6055, operating up to 550 °C, uses molten salts with a pumping system for maintaining the necessary consistent fluid level required for liquid-in-glass thermometer calibrations. A viewing channel is built into the top cover for a clear visual path to your glass thermometers.
The 6055 also has an optional thermometer carousel for holding several glass thermometers in the correct calibration position without exposing them to the hot salts in the bath. The Model 2018 Carousel is completely constructed of stainless steel and has an elevated handle for rotating your thermometers to the viewing position.
These deep-well baths are built to the same performance standards as all Fluke Calibration baths, which means you can’t find another bath that has better stability or uniformity.
|
7007 |
6054 |
6055 |
Range |
–5 °C to 110 °C | 50 °C to 300 °C | 200 °C to 550 °C |
Temperature Stability | ± 0.001 °C at 0 °C (water/ethanol) ± 0.001 °C at 25 °C (water) ± 0.003 °C at 100 °C (oil 5012) |
± 0.003 °C at 100 °C (oil 5012) ± 0.005 °C at 300 °C (oil 5017) |
± 0.003 °C at 200 °C (salt) ± 0.01 °C at 550 °C (salt) |
Temperature Uniformity | ± 0.004 °C at 0 °C (ethanol) ± 0.007 °C at 100 °C (oil 5012) |
± 0.007 °C at 100 °C (oil 5012) ± 0.015 °C at 300 °C (oil 5017) |
± 0.005 °C at 200 °C (salt) ± 0.010 °C at 550 °C (salt) |
Temperature Setting |
Digital display with push-button data entry | ||
Set-Point Resolution |
0.002 °C, high res. 0.00003 °C | 0.01 °C, high res. 0.00018 °C | |
Display Temperature Resolution |
0.01 °C | ||
Digital Setting Accuracy |
± 1 °C | ||
Digital Setting Repeatability |
± 0.005 °C | ± 0.01 °C | |
Heaters |
250 to 1000 W | 250 to 1000 W | 260 to 2080 W |
Working Area |
178 mm dia. (7 in) | 196 mm dia. (7.7 in) | 107 mm dia. (4.2 in) |
Depth |
24 in | 24 in | 24 in |
Wetted Parts |
304 stainless steel | ||
Power | 230 V ac (± 10 %), 50 or 60 Hz, 15 A (Specify frequency, contact Fluke Calibration if CE mark required.) |
230 V ac (± 10 %), 50/60 Hz, 15 A max |
230 V ac (± 10 %), 50/60 Hz, 15 A |
Volume | 42 liters (11.2 gal) | 50 liters (13.2 gal) | 19.8 liters (5.2 gal, 43 kg [95 lb] of bath salt) |
Size (D x W x H) | 470 x 775 x 194 mm to working surface (18.5 x 30.5 x 47 in), 1397 mm (55 in) to top of stir motor, 914 mm (36 in) to control panel | 572 x 762 x 1219 mm to working surface (22.5 x 30 x 48 in), 1422 mm (56 in) to top of stir motor box, 914 mm (36 in) to control panel | 572 x 775 x 1219 mm to working surface (22.5 x 30.5 x 48 in), 1524 mm (60 in) to top of stir motor box, 914 mm (36 in) to control panel |
Distance from Line of Sight to Top of Fluid |
9.5 mm (3/8 in) | 15.9 mm (5/8 in) | |
Automation | Use the 1586A Super-DAQ to automate temperature sensor calibration. For IEEE-488, add the 2001-IEEE to the automation package. | ||
Weight |
70.8 kg (156 lb) |
製品マニュアル |
---|
アプリケーション・ノート |
---|
論文/アプリケーション・ノート |
---|
Fluke Calibration Bath Safety Features |
How to Choose a Calibration Bath |
On-Demand Webinars |
---|
How to Select the Right Temperature Calibration Bath |
- Home
- Products
- New Products
- Electrical Calibration
- RF Calibration
- Data Acquisition and Test Equipment
- Temperature Calibration
- Humidity Calibration
- Pressure Calibration
- Flow Calibration
- Process Calibration Tools
- Calibration Software
- Service and Support
- All Calibration Instruments
- Handheld Test Tools
- Purchase Info
- News
- Training and Events
- Literature and Education
- Service and Support
- About Us
Sidebar Request a Quote
見積依頼フォーム